抄録
The first experimental evidence of three-dimensional conformai irradiation is performed by using a new type of filter developed for charged particle radiotherapy. The new filter can yield a static irradiation field where the width of the spread-out Bragg peak is adjusted to the target as a two-dimensional continuous function in the transverse plane. The filter is made of many layers produced by using stereolithography. In the filter, a structure with two regions with different shaped miniaturized ridges is adopted to get a flat dose profile on the spread-out peak. A conformal field around a spherical target is realized by a filter that has a shape corresponding to the outward form of the target.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 234-236 |
| ページ数 | 3 |
| ジャーナル | Review of Scientific Instruments |
| 巻 | 72 |
| 号 | 1 I |
| DOI | |
| 出版ステータス | 出版済み - 1月 2001 |
| 外部発表 | はい |