抄録
The development of a multielectrode array is the key issue for retinal prostheses. We developed a 10 × 10 platinum electrode array that consists of an 8-μm polyimide layer sandwiched between 5-μm polymonochloro-para- xylylene (parylene-C) layers. Each electrode was formed as a 30-μm-high bump by Pt/Au double-layer electroplating. We estimated the charge delivery capability (CDC) of the electrode by measuring the CDCs of two-channel electrode arrays. The dimensions of each electrode of the two-channel array were the same as those of each electrode formed on the 10 × 10 array. The results suggest that for cathodic-first (CF) pulses, 80% of electrodes surpassed our development target of 318 μC/cm2, which corresponds to the charge density of pulses of 500 μs duration and 200 μA amplitude for a 200-μm-diameter planar electrode.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 263-266 |
| ページ数 | 4 |
| ジャーナル | Journal of Artificial Organs |
| 巻 | 9 |
| 号 | 4 |
| DOI | |
| 出版ステータス | 出版済み - 12月 2006 |