抄録
A stacked NaI(Tl) spectrometer which consists of two silicon detectors, 4 rectangular NaI(Tl) scintillators (2 inches ×2 inches ×2 inches) and a cylindrical NaI(Tl) scintillator (φ8 cm × 18 cm) has been developed to detect spectrum of charged particles up to about 400MeV. Tests were performed using 300 and 350MeV proton beams from the ring cyclotron at the Research Center for Nuclear Physics (RCNP), Osaka university. The ΔE-E technique was used to identify particles and also events which originated to nuclear reactions in the detector materials. The tail of peak spectrum is suppressed clearly and the peak efficiency is reasonably estimated.
本文言語 | 英語 |
---|---|
ページ | 800-804 |
ページ数 | 5 |
出版ステータス | 出版済み - 1995 |
イベント | Proceedings of the 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 3) - San Francisco, CA, USA 継続期間: 10月 21 1995 → 10月 28 1995 |
その他
その他 | Proceedings of the 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference. Part 1 (of 3) |
---|---|
City | San Francisco, CA, USA |
Period | 10/21/95 → 10/28/95 |
!!!All Science Journal Classification (ASJC) codes
- 核物理学および高エネルギー物理学
- 放射線
- 放射線学、核医学およびイメージング