Realization of matching conditions for high-resolution spectrometers

H. Fujita, Y. Fujita, G. P.A. Berg, A. D. Bacher, C. C. Foster, K. Hara, K. Hatanaka, T. Kawabata, T. Noro, H. Sakaguchi, Y. Shimbara, T. Shinada, E. J. Stephenson, H. Ueno, M. Yosoi

研究成果: ジャーナルへの寄稿学術誌査読

122 被引用数 (Scopus)

抄録

For precise measurements of nuclear-reaction spectra using a beam from an accelerator, a high-resolution magnetic spectrometer is a powerful tool. The full capability of a magnetic spectrometer, however, can be achieved only if the characteristics of the beam coming from the accelerator are matched to those required by the spectrometer by properly adjusting the beam line conditions. The matching methods, lateral dispersion matching, focus matching and also the kinematic correction compensate the spectrum line-broadening effects caused by the beam momentum spread and reaction kinematics. In addition, angular dispersion matching should be performed if good resolution of the scattering angle is important. Diagnostic methods developed to realize these matching conditions for the spectrometers K600 at IUCF and Grand Raiden at RCNP are presented.

本文言語英語
ページ(範囲)17-26
ページ数10
ジャーナルNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
484
1-3
DOI
出版ステータス出版済み - 5月 21 2002

!!!All Science Journal Classification (ASJC) codes

  • 核物理学および高エネルギー物理学