Performance of TES X-ray microcalorimeters with a novel absorber design

Simon R. Bandler, R. P. Brekosky, A. D. Brown, J. A. Chervenak, E. Figueroa-Feliciano, F. M. Finkbeiner, N. Iyomoto, R. L. Kelley, C. A. Kilbourne, F. S. Porter, J. Sadleir, S. J. Smith

研究成果: ジャーナルへの寄稿学術誌査読

72 被引用数 (Scopus)

抄録

Superconducting transition-edge sensor (TES) microcalorimeters have demonstrated the Constellation-X requirements for spectral resolution, speed, and pixel size in a close-packed geometry. We will present our recent breakthrough energy resolution with sensors that have all gold and bismuth-gold absorbers. This has been enabled by cantilevered absorbers that make contact to the TES only in regions that are not part of the active thermometer. With this approach, rapid thermalization of the x-ray energy is achieved and interaction between the absorber and TES sensor films is avoided. This design allows us to obtain uniform high performance and is compatible with large-format, high fill-factor arrays. We will discuss this design, the results we have achieved in 8×8 arrays of these pixels, and the dependence of the performance on the geometry of the absorber contact area and on stress within the sensor.

本文言語英語
ページ(範囲)400-405
ページ数6
ジャーナルJournal of Low Temperature Physics
151
1-2 PART 1
DOI
出版ステータス出版済み - 4月 1 2008
外部発表はい

!!!All Science Journal Classification (ASJC) codes

  • 原子分子物理学および光学
  • 材料科学(全般)
  • 凝縮系物理学

フィンガープリント

「Performance of TES X-ray microcalorimeters with a novel absorber design」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。