Investigation of optimum grid for digital radiography

Kentaro Naka, Nobukazu Tanaka, Seiji Kumazawa, Junji Morishita, Hidetaka Arimura, Fukai Toyofuku, Masafumi Ohki, Yoshiharu Higashida

研究成果: 書籍/レポート タイプへの寄稿会議への寄与

抄録

The purpose of this study was to investigate the optimal anti-scatter grid (grid) selection for digital radiography with a computed radiography (CR) system. Anterior-posterior lumbar spine radiographies were performed with no grid and grid ratios of 5, 8, 10, 12 and 14 under the same exposure conditions. An x-ray tube voltage of 80 kV and 20 mAs were used in all radiographies. Signal-to-noise-ratio (SNR) for all radiographic technique was measured as an image quality index. To consider the effect of image quality, scatter fractions, incident doses to imaging plate (IP) were also measured. The use of grid ratios of 10, 12 and 14 provided higher SNR than those of the images with grid ratios of 5 and 8 when same exposure conditions were used. Phantom radiographs also indicated that the images with 10, 12 and 14 grid ratios improved image quality over that obtainable with images of lower grid ratios although the images look noisy.

本文言語英語
ホスト出版物のタイトルWorld Congress on Medical Physics and Biomedical Engineering
ホスト出版物のサブタイトルDiagnostic Imaging
出版社Springer Verlag
ページ149-151
ページ数3
2
ISBN(印刷版)9783642038785
DOI
出版ステータス出版済み - 2009
イベントWorld Congress on Medical Physics and Biomedical Engineering: Diagnostic Imaging - Munich, ドイツ
継続期間: 9月 7 20099月 12 2009

出版物シリーズ

名前IFMBE Proceedings
番号2
25
ISSN(印刷版)1680-0737

その他

その他World Congress on Medical Physics and Biomedical Engineering: Diagnostic Imaging
国/地域ドイツ
CityMunich
Period9/7/099/12/09

!!!All Science Journal Classification (ASJC) codes

  • バイオエンジニアリング
  • 生体医工学

フィンガープリント

「Investigation of optimum grid for digital radiography」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル