Illumination semiangle of 10-9 rad achieved in a 1.2-MV atomic resolution holography transmission electron microscope

Tetsuya Akashi, Yoshio Takahashi, Ken Harada, Toshio Onai, Yoshimasa A. Ono, Hiroyuki Shinada, Yasukazu Murakami

研究成果: ジャーナルへの寄稿レター査読

4 被引用数 (Scopus)

抄録

The coherency of a 1.2-MV transmission electron microscope was evaluated through illumination semiangles calculated from lengths over which Fresnel fringes can be observed. These lengths were determined from the diameters of circular holes fully filled with Fresnel fringes, i.e. this method allows lengths to be accurately measured even if micrographs are subjected to distortions. The smallest illumination semiangle of 4.0 × 10-9 rad was obtained for a circular hole with a diameter of 191 μm. In addition, electron beam brightness was estimated to be approximately 3 × 1014 A/m2·sr from the obtained illumination semiangle values and current densities. The results provide us with essential information that can be referred to in future electron holography studies aimed at detecting weak electromagnetic fields in materials.

本文言語英語
ページ(範囲)286-290
ページ数5
ジャーナルMicroscopy
67
5
DOI
出版ステータス出版済み - 10月 1 2018

!!!All Science Journal Classification (ASJC) codes

  • 構造生物学
  • 器械工学
  • 放射線学、核医学およびイメージング

フィンガープリント

「Illumination semiangle of 10-9 rad achieved in a 1.2-MV atomic resolution holography transmission electron microscope」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。