Harmonization of Standardized Uptake Value among Different Generation PET/ CT Cameras Based on a Phantom Experiment -Utility of SUV(peak)

Go Akamatsu, Hiroyuki Nishida, Akira Fujino, Akihito Ohnishi, Yasuhiko Ikari, Tomoyuki Nishio, Akira Maebatake, Masayuki Sasaki, Michio Senda

研究成果: ジャーナルへの寄稿学術誌査読

2 被引用数 (Scopus)

抄録

Standardized uptake value (SUV) has been widely used as a semi-quantitative metric of uptake in FDGPET/ CT for diagnosis of malignant tumors and evaluation of tumor therapies. However, the SUV depends on various factors including PET/CT scanner specifications and reconstruction parameters. The purpose of this study is to harmonize the SUV among two PET/CT models of different generation: two units of Discovery ST Elite Performance(DSTEP) and Discovery 690 (D690) PET/CT scanners. The NEMA body phantom filled with 18F solution was scanned for 30 minutes in list-mode. The D690 PET images were reconstructed with OSEM, OSEM+TOF, and OSEM+PSF. Gaussian post-filters of 4-9 mm FWHM were applied to find the parameters that provides harmonized SUV. We determined the SUV-harmonized parameter for each reconstruction algorithm. Then, the 10 PET images simulating clinical scan conditions were respectively generated to evaluate the bias and variability of SUV(max) and SUV(peak). The SUV(max) strongly depended not only on spatial resolution but also on image noise. On the other hand, the SUV(peak) was a robust metric to image noise level. TOF improved the variability of SUV(max) and SUV(peak). Thus, we were able to harmonize the spatial resolution using SUV(peak) based on the phantom study. Because SUV(max) was also strongly affected by image noise, sufficient count statistics is essential for SUV(max) harmonization. We recommended that TOF reconstruction and SUV(peak) metric should be used to harmonize SUV.

本文言語英語
ページ(範囲)735-745
ページ数11
ジャーナルNihon Hōshasen Gijutsu Gakkai zasshi
71
9
DOI
出版ステータス出版済み - 9月 1 2015
外部発表はい

!!!All Science Journal Classification (ASJC) codes

  • 医学(全般)

フィンガープリント

「Harmonization of Standardized Uptake Value among Different Generation PET/ CT Cameras Based on a Phantom Experiment -Utility of SUV(peak)」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル