We developed a lost alpha detection system to use in burning plasma experiments. The scintillators of Ag:ZnS and polycrystalline Ce:YAG were designed for a high-temperature environment, and the optical transmission line was designed to transmit from the scintillator to the port plug. The required optical components of lenses and mirrors were irradiated using the fission reactor with the initial result that there was no clear change after the irradiation with a neutron flux of 9.6× 1017 nm-2 s-1 for 48 h. We propose a diagnostic of alpha particle loss, so-called alpha particle induced gamma ray spectroscopy. The initial laboratory test has been carried out by the use of the Ce doped Lu2SiO 5 scintillator detector and an Am-Be source to detect the 4.44 MeV high energy gamma ray due to the 9Be (α,nγ) 12C reaction.

本文言語英語
論文番号10D313
ジャーナルReview of Scientific Instruments
81
10
DOI
出版ステータス出版済み - 10月 2010
外部発表はい