抄録
The pulse shape discrimination (PSD) is often made for liquid organic scintillators for the purpose of separating neutrons from gamma-rays. Particles emitted from spallation reactions induced by protons of GeV range were measured with an NE213 liquid scintillator, 12.7 cm in diameter and 12.7 cm thick. The pulse shape characteristics were studied by the use of charged-particles incident on the detector. A two-gate integration method. The charged particles (protons and pions) having velocities below β=0.133 were identified into the conventional recoil-proton like events. On the contrary, the charged particles aboveβ=0.133 were classified as the electron-like events in the PSD. The PSD was useful only for neutrons which produce recoil protons stopping within the scintillator. The experimental results suggest that a great care should be taken in the PSD for separation of intermediate-energy neutrons from gamma-rays.
| 本文言語 | 英語 |
|---|---|
| ページ | 6/219-6/221 |
| 出版ステータス | 出版済み - 2000 |
| イベント | 2000 IEEE Nuclear Science Symposium Conference Record - Lyon, フランス 継続期間: 10月 15 2000 → 10月 20 2000 |
その他
| その他 | 2000 IEEE Nuclear Science Symposium Conference Record |
|---|---|
| 国/地域 | フランス |
| City | Lyon |
| Period | 10/15/00 → 10/20/00 |
!!!All Science Journal Classification (ASJC) codes
- 核物理学および高エネルギー物理学
- 放射線
- 放射線学、核医学およびイメージング
フィンガープリント
「Deterioration of pulse-shape discrimination in liquid organic scintillator at high energies」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS