抄録
Monoenergetic X-rays extracted from synchrotron radiation were detected by two kinds of Nb-based superconducting tunnel junctions. Each of junctions had been fabricated by using different fabrication processes. Measurements were carried out under severe noisy circumstances. A cooled-FET preamplifier was employed to improve the signal to noise ratio. Pulse height distributions of the output signal of the preamplifier were obtained for monoenergetic X-rays in a range of 4 to 15 keV. Values of the energy for single electron collection were found to exhibit a nonlinear dependence on the energy of incident X-rays.
本文言語 | 英語 |
---|---|
ページ(範囲) | 29-34 |
ページ数 | 6 |
ジャーナル | Denshi Gijutsu Sogo Kenkyusho Iho/Bulletin of the Electrotechnical Laboratory |
巻 | 61 |
号 | 6-7 |
出版ステータス | 出版済み - 12月 1 1997 |
!!!All Science Journal Classification (ASJC) codes
- 凝縮系物理学