抄録
Wearable wireless physiological sensors are helpful for monitoring and maintaining human health. Blood flow contains abundant physiological information but it is hard to measure blood flow during exercise using conventional blood flowmeters because of their size, weight, and use of optic fibers. To resolve these disadvantages, we previously developed a micro integrated laser Doppler blood flowmeter using microelectromechanical systems technology. This micro blood flowmeter is wearable and capable of stable measurement signals even during movement. Therefore, we attempted to measure skin blood flow at the forehead, fingertip, and earlobe of seven young men while running as a pilot experiment to extend the utility of the micro blood flowmeter. We measured blood flow in each subject at velocities of 6, 8, and 10 km/h. We succeeded in obtaining stable measurements of blood flow, with few motion artifacts, using the micro blood flowmeter, and the pulse wave signal and motion artifacts were clearly separated by conducting frequency analysis. Furthermore, the results showed that the extent of the changes in blood flow depended on the intensity of exercise as well as previous work with an ergometer. Thus, we demonstrated the capability of this wearable blood flow sensor for measurement during exercise.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 25507-25519 |
| ページ数 | 13 |
| ジャーナル | Sensors (Switzerland) |
| 巻 | 15 |
| 号 | 10 |
| DOI | |
| 出版ステータス | 出版済み - 10月 5 2015 |
UN SDG
この成果は、次の持続可能な開発目標に貢献しています
-
SDG 3 すべての人に健康と福祉を
!!!All Science Journal Classification (ASJC) codes
- 分析化学
- 情報システム
- 原子分子物理学および光学
- 生化学
- 器械工学
- 電子工学および電気工学
フィンガープリント
「Detection of site-specific blood flow variation in humans during running by a wearable laser Doppler flowmeter」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS