TY - JOUR
T1 - Speed modulation of the continuous-flow total artificial heart to simulate a physiologic arterial pressure waveform
AU - Shiose, Akira
AU - Nowak, Kathleen
AU - Horvath, David J.
AU - Massiello, Alex L.
AU - Golding, Leonard A.R.
AU - Fukamachi, Kiyotaka
PY - 2010/9
Y1 - 2010/9
N2 - This study demonstrated the concept of using speed modulation in a continuous-flow total artificial heart (CFTAH) to shape arterial pressure waveforms and to adjust pressure pulsatility. A programmable function generator was used to determine the optimum pulsatile speed profile. Three speed profiles [sinusoidal, rectangular, and optimized (a profile optimized for generation of a physiologic arterial pressure waveform)] were evaluated using the CFTAH mock circulatory loop. Hemodynamic parameters were recorded at average pump speeds of 2,700 rpm and a modulation cycle of 60 beats per minute. The effects of varying physiologically relevant vascular resistance and lumped compliance on the hemodynamics were assessed. The feasibility of using speed modulation to manipulate systemic arterial pressure waveforms, including a physiologic pressure waveform, was demonstrated in vitro. The additional pump power consumption needed to generate a physiologic pulsatile pressure was 16.2% of the power consumption in nonpulsatile continuous-flow mode. The induced pressure waveforms and pulse pressure were shown to be very responsive to changes in both systemic vascular resistance and arterial compliance. This system also allowed pulsatile pulmonary arterial waveform. Speed modulation in the CFTAH could enable physicians to obtain desired pressure waveforms by simple manual adjustment of speed control input waveforms.
AB - This study demonstrated the concept of using speed modulation in a continuous-flow total artificial heart (CFTAH) to shape arterial pressure waveforms and to adjust pressure pulsatility. A programmable function generator was used to determine the optimum pulsatile speed profile. Three speed profiles [sinusoidal, rectangular, and optimized (a profile optimized for generation of a physiologic arterial pressure waveform)] were evaluated using the CFTAH mock circulatory loop. Hemodynamic parameters were recorded at average pump speeds of 2,700 rpm and a modulation cycle of 60 beats per minute. The effects of varying physiologically relevant vascular resistance and lumped compliance on the hemodynamics were assessed. The feasibility of using speed modulation to manipulate systemic arterial pressure waveforms, including a physiologic pressure waveform, was demonstrated in vitro. The additional pump power consumption needed to generate a physiologic pulsatile pressure was 16.2% of the power consumption in nonpulsatile continuous-flow mode. The induced pressure waveforms and pulse pressure were shown to be very responsive to changes in both systemic vascular resistance and arterial compliance. This system also allowed pulsatile pulmonary arterial waveform. Speed modulation in the CFTAH could enable physicians to obtain desired pressure waveforms by simple manual adjustment of speed control input waveforms.
UR - http://www.scopus.com/inward/record.url?scp=77956394521&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77956394521&partnerID=8YFLogxK
U2 - 10.1097/MAT.0b013e3181e650f8
DO - 10.1097/MAT.0b013e3181e650f8
M3 - Article
C2 - 20616704
AN - SCOPUS:77956394521
SN - 1058-2916
VL - 56
SP - 403
EP - 409
JO - ASAIO Journal
JF - ASAIO Journal
IS - 5