Abstract
Coprecipitation with gallium hydroxide is studied for the preconcentration of trace metals in sea water before multi-element analysis by inductively-coupled plasma/atomic emission spectrometry. Gallium precipitates at pH 9 only when magnesium is present. Optimum conditions are established for multi-element preconcentration and removal of matrix elements. The method is almost free from contamination because of the use of highly pure gallium meetal and only a small amount of sodium hydroxide for pH adjustment. Spectral interferences from gallium are negligible and a concentratioin factor of more than 200 can be obtained. Detection limits range from a few ng l-1 to 150 ng l- for Al, Co, Cr, Fe, La, Mn, Ni, Ti, V, Zn, Y and Pb. Artificial and natural sea-water samples can be analyzed with adequate precision.
| Original language | English |
|---|---|
| Pages (from-to) | 139-151 |
| Number of pages | 13 |
| Journal | Analytica Chimica Acta |
| Volume | 177 |
| Issue number | C |
| DOIs | |
| Publication status | Published - 1985 |
| Externally published | Yes |
All Science Journal Classification (ASJC) codes
- Analytical Chemistry
- Biochemistry
- Environmental Chemistry
- Spectroscopy
Fingerprint
Dive into the research topics of 'Simultaneous multi-element determination of trace metals in sea water by inductively-coupled plasma atomic emission spectrometry after coprecipitation with gallium'. Together they form a unique fingerprint.Cite this