TY - JOUR
T1 - Left Ventricular Mechanical Unloading by Total Support of Impella in Myocardial Infarction Reduces Infarct Size, Preserves Left Ventricular Function, and Prevents Subsequent Heart Failure in Dogs
AU - Saku, Keita
AU - Kakino, Takamori
AU - Arimura, Takahiro
AU - Sunagawa, Genya
AU - Nishikawa, Takuya
AU - Sakamoto, Takafumi
AU - Kishi, Takuya
AU - Tsutsui, Hiroyuki
AU - Sunagawa, Kenji
N1 - Publisher Copyright:
© 2018 American Heart Association, Inc.
PY - 2018/5/1
Y1 - 2018/5/1
N2 - BACKGROUND: Acute myocardial infarction remains a leading cause of chronic heart failure. Excessive myocardial oxygen demand relative to supply is the fundamental mechanism of myocardial infarction. We thus hypothesized that left ventricular (LV) mechanical unloading by the total support of transvascular LV assist device Impella could minimize oxygen demand, thereby reducing infarct size and preventing subsequent heart failure. METHODS AND RESULTS: In 20 dogs, we ligated the left anterior descending coronary artery for 180 minutes and then reperfused. We introduced Impella from 60 minutes after the onset of ischemia to 60 minutes after reperfusion. In the partial support group, Impella supported 50% of total cardiac output. In the total support group, systemic flow totally depends on Impella flow. Four weeks after ischemia/reperfusion (I/R), we compared LV function and infarct size among 4 groups: sham (no I/R), I/R (no Impella support), partial support, and total support. Compared with I/R, total support lowered LV end-diastolic pressure (15.0±3.5 versus 4.7±1.7 mm Hg; P<0.001), increased LV end-systolic elastance (4.3±0.8 versus 13.9±5.1 mm Hg/mL; P<0.001), and decreased NT-proBNP (N-terminal pro-B-type natriuretic peptide) level (4081±1123 versus 1773±390 pg/mL; P<0.05). Furthermore, total support markedly reduced infarct size relative to I/R, whereas partial support decreased infarct size to a lesser extent (I/R, 16.3±2.6; partial support, 8.5±4.3; and total support, 2.1±1.6%; P<0.001). CONCLUSIONS: LV mechanical unloading by the total support of Impella during the acute phase of myocardial infarction reduced infarct size and prevented subsequent heart failure in dogs.
AB - BACKGROUND: Acute myocardial infarction remains a leading cause of chronic heart failure. Excessive myocardial oxygen demand relative to supply is the fundamental mechanism of myocardial infarction. We thus hypothesized that left ventricular (LV) mechanical unloading by the total support of transvascular LV assist device Impella could minimize oxygen demand, thereby reducing infarct size and preventing subsequent heart failure. METHODS AND RESULTS: In 20 dogs, we ligated the left anterior descending coronary artery for 180 minutes and then reperfused. We introduced Impella from 60 minutes after the onset of ischemia to 60 minutes after reperfusion. In the partial support group, Impella supported 50% of total cardiac output. In the total support group, systemic flow totally depends on Impella flow. Four weeks after ischemia/reperfusion (I/R), we compared LV function and infarct size among 4 groups: sham (no I/R), I/R (no Impella support), partial support, and total support. Compared with I/R, total support lowered LV end-diastolic pressure (15.0±3.5 versus 4.7±1.7 mm Hg; P<0.001), increased LV end-systolic elastance (4.3±0.8 versus 13.9±5.1 mm Hg/mL; P<0.001), and decreased NT-proBNP (N-terminal pro-B-type natriuretic peptide) level (4081±1123 versus 1773±390 pg/mL; P<0.05). Furthermore, total support markedly reduced infarct size relative to I/R, whereas partial support decreased infarct size to a lesser extent (I/R, 16.3±2.6; partial support, 8.5±4.3; and total support, 2.1±1.6%; P<0.001). CONCLUSIONS: LV mechanical unloading by the total support of Impella during the acute phase of myocardial infarction reduced infarct size and prevented subsequent heart failure in dogs.
UR - http://www.scopus.com/inward/record.url?scp=85059473676&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85059473676&partnerID=8YFLogxK
U2 - 10.1161/CIRCHEARTFAILURE.117.004397
DO - 10.1161/CIRCHEARTFAILURE.117.004397
M3 - Article
C2 - 29739745
AN - SCOPUS:85059473676
SN - 1941-3297
VL - 11
SP - e004397
JO - Circulation. Heart failure
JF - Circulation. Heart failure
IS - 5
ER -