T1 - Highly sensitive third-harmonic detection method of magnetic nanoparticles using an AC susceptibility measurement system for liquid-phase assay
AU - Mizoguchi, Takako
AU - Kandori, Akihiko
AU - Kawabata, Ryuzo
AU - Ogata, Kuniomi
AU - Hato, Tsunehiro
AU - Tsukamoto, Akira
AU - Adachi, Seiji
AU - Tanabe, Keiichi
AU - Tanaka, Saburo
AU - Tsukada, Keiji
AU - Enpuku, Keiji
N1 - Funding Information:
This work was supported by the Strategic Promotion of Innovative Research and Development Program through the Japan Science and Technology Agency. This paper was recommended by Editor A. Kleinsasser.
Publisher Copyright:
© 2002-2011 IEEE.
PY - 2016/8
Y1 - 2016/8
N2 - Magnetic sensing techniques have been recently developed to detect biomarkers using magnetic nanoparticles (MNPs) in liquid-phase assays. In the case of alternating current (ac) magnetic susceptibility when detecting a low number of MNPs, the diamagnetic signal of water becomes a major problem at a low signal-to-noise ratio. Therefore, we developed a high-temperature superconducting-quantum-interference-device-based ac susceptibility measurement system that can detect third-harmonic signals from MNPs. On the basis of the nonlinear characteristic of MNPs with large ac magnetic excitation (1.06 kHz), the third-harmonic signal detection (3.18 kHz) of MNPs leads to the elimination of the diamagnetic signal of water. The system consists of excitation and gradiometer pickup coils and a multifunctional circuit device. To improve the signal-to-noise ratio of the third-harmonic measurement, the system has two cancelation functions regarding the fundamental magnetic field. The MNPs are magnetized by using the fundamental magnetic field using the excitation coil. Third-harmonic signals from the MNPs were then detected during the up-and-down movement of the MNPs. We evaluated the detection sensitivity of our system using MNPs. We confirmed that the limit of detection consistency of MNPs was 10 ng by using the third-harmonic measurement method.
AB - Magnetic sensing techniques have been recently developed to detect biomarkers using magnetic nanoparticles (MNPs) in liquid-phase assays. In the case of alternating current (ac) magnetic susceptibility when detecting a low number of MNPs, the diamagnetic signal of water becomes a major problem at a low signal-to-noise ratio. Therefore, we developed a high-temperature superconducting-quantum-interference-device-based ac susceptibility measurement system that can detect third-harmonic signals from MNPs. On the basis of the nonlinear characteristic of MNPs with large ac magnetic excitation (1.06 kHz), the third-harmonic signal detection (3.18 kHz) of MNPs leads to the elimination of the diamagnetic signal of water. The system consists of excitation and gradiometer pickup coils and a multifunctional circuit device. To improve the signal-to-noise ratio of the third-harmonic measurement, the system has two cancelation functions regarding the fundamental magnetic field. The MNPs are magnetized by using the fundamental magnetic field using the excitation coil. Third-harmonic signals from the MNPs were then detected during the up-and-down movement of the MNPs. We evaluated the detection sensitivity of our system using MNPs. We confirmed that the limit of detection consistency of MNPs was 10 ng by using the third-harmonic measurement method.
UR - http://www.scopus.com/inward/record.url?scp=84979633292&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84979633292&partnerID=8YFLogxK
U2 - 10.1109/TASC.2016.2581703
DO - 10.1109/TASC.2016.2581703
M3 - Article
AN - SCOPUS:84979633292
SN - 1051-8223
VL - 26
JO - IEEE Transactions on Applied Superconductivity
JF - IEEE Transactions on Applied Superconductivity
IS - 5
M1 - 7492213
ER -