Establishment of mail medical examination system using immediate plasma separating device by the self-collection blood--the method of dilution ratio calculation by using Internal standard for the sample with different amount of collecting blood

Masatoshi Horita, Shinya Sugimoto, Eisaku Hokazono, Susumu Osawa

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

A method of calibrating unquantified samples was developed in order to conduct tests from approximately 65 microl or less of self-collected whole blood from the finger. In this method, the dilution ratio is estimated by absorption analysis of an internal standard substance in the blood dilution buffer. Good results for replication of the dilution ratio were obtained, with a CV value of 0.99-2.78% and a rate of concordance with the theoretical dilution ratio of 99.4-99.8%. Evaluations of the utility of this system (Demecal System) were performed using total cholesterol measurement, and excellent correlation with the measurement value in plasma was achieved, with a correlation coefficient of r=0.975 at n=40 and a regression equation of y=1.02x-2.83. Dilution of 7-fold or less is required to assure accuracy with this system.

Original languageEnglish
Pages (from-to)577-583
Number of pages7
JournalRinsho byori. The Japanese journal of clinical pathology
Volume56
Issue number7
Publication statusPublished - Jul 2008
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • General Medicine

Fingerprint

Dive into the research topics of 'Establishment of mail medical examination system using immediate plasma separating device by the self-collection blood--the method of dilution ratio calculation by using Internal standard for the sample with different amount of collecting blood'. Together they form a unique fingerprint.

Cite this