TY - JOUR
T1 - Dose evaluation indices for total body irradiation using TomoDirect with different numbers of ports
T2 - A comparison with the TomoHelical method
AU - Kasai, Yuki
AU - Fukuyama, Yukihide
AU - Terashima, Hiromi
AU - Nakamura, Katsumasa
AU - Sasaki, Tomonari
N1 - Funding Information:
The authors would like to thank Tadashi Nakabayashi (Accuray Japan K.K.) for useful comments. This work was supported by JSPS KAKENHI Grant Number JP17K10483.
Publisher Copyright:
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
PY - 2019/2/1
Y1 - 2019/2/1
N2 - TomoDirect has been reported to have some advantages over TomoHelical in delivering total body irradiation (TBI). This study aimed to investigate the relationships between the number of ports and the dose evaluation indices in low-dose TBI in TomoDirect mode using 2–12 ports and to compare these data with those for the TomoHelical mode in a simulation study. Thirteen patients underwent low-dose TBI in TomoHelical mode from June 2015 to June 2016. We used the same computed tomography data sets for these patients to create new treatment plans for upper-body parts using TomoDirect mode with 2–12 beam angles as well as TomoHelical mode. The prescription was 4 Gy in two equal fractions. For the TomoDirect data, we generated plans with 2–12 ports with approximately equally spaced angles; the modulation factor, field width, and pitch were 2.0, 5.0 cm, and 0.500, respectively. For the TomoHelical plans, the modulation factor, field width, and pitch were 2.0, 5.0 cm, and 0.397, respectively. D2, D98, D50, and the homogeneity index (HI) were evaluated to compare TomoDirect plans having 2–12 ports with the TomoHelical plan. Using TomoDirect plans, D2 with four ports or fewer, D98 with 10 ports or fewer, D50 with four ports or fewer and HI with five ports or fewer showed statistically significantly worse results than the TomoHelical plan. With the TomoDirect plans, D2 with seven ports or more, D50 with eight ports or more, and HI with eight ports or more showed statistically significant improvement compared with the TomoHelical plan. All of the dose evaluation indices of the TomoDirect plans showed a tendency to improve as the number of ports increased. TomoDirect plans showed statistically significant improvement of D2, D50, and HI compared with the TomoHelical plan. Therefore, we conclude that TomoDirect can provide better dose distribution in low-dose TBI with TomoTherapy.
AB - TomoDirect has been reported to have some advantages over TomoHelical in delivering total body irradiation (TBI). This study aimed to investigate the relationships between the number of ports and the dose evaluation indices in low-dose TBI in TomoDirect mode using 2–12 ports and to compare these data with those for the TomoHelical mode in a simulation study. Thirteen patients underwent low-dose TBI in TomoHelical mode from June 2015 to June 2016. We used the same computed tomography data sets for these patients to create new treatment plans for upper-body parts using TomoDirect mode with 2–12 beam angles as well as TomoHelical mode. The prescription was 4 Gy in two equal fractions. For the TomoDirect data, we generated plans with 2–12 ports with approximately equally spaced angles; the modulation factor, field width, and pitch were 2.0, 5.0 cm, and 0.500, respectively. For the TomoHelical plans, the modulation factor, field width, and pitch were 2.0, 5.0 cm, and 0.397, respectively. D2, D98, D50, and the homogeneity index (HI) were evaluated to compare TomoDirect plans having 2–12 ports with the TomoHelical plan. Using TomoDirect plans, D2 with four ports or fewer, D98 with 10 ports or fewer, D50 with four ports or fewer and HI with five ports or fewer showed statistically significantly worse results than the TomoHelical plan. With the TomoDirect plans, D2 with seven ports or more, D50 with eight ports or more, and HI with eight ports or more showed statistically significant improvement compared with the TomoHelical plan. All of the dose evaluation indices of the TomoDirect plans showed a tendency to improve as the number of ports increased. TomoDirect plans showed statistically significant improvement of D2, D50, and HI compared with the TomoHelical plan. Therefore, we conclude that TomoDirect can provide better dose distribution in low-dose TBI with TomoTherapy.
UR - http://www.scopus.com/inward/record.url?scp=85061198706&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85061198706&partnerID=8YFLogxK
U2 - 10.1002/acm2.12540
DO - 10.1002/acm2.12540
M3 - Article
C2 - 30721567
AN - SCOPUS:85061198706
SN - 1526-9914
VL - 20
SP - 129
EP - 135
JO - Journal of Applied Clinical Medical Physics
JF - Journal of Applied Clinical Medical Physics
IS - 2
ER -