T1 - Determination of cyanide and thiocyanate in blood by gas chromatography and gas chromatography-mass spectrometry
AU - Kage, Shigetoshi
AU - Nagata, Takeaki
AU - Kudo, Keiko
PY - 1996/1/12
Y1 - 1996/1/12
N2 - We devised a sensitive and simple method for determining cyanide and its major metabolite, thiocyanate, in blood using an extractive alkylation technique. Pentafluorobenzyl bromide was used as the alkylating agent, and tetradecyldimethylbenzylammonium chloride was used as the phase-transfer catalyst. The derivatives obtained were analyzed qualitatively by gas chromatography-mass spectrometry and quantitatively by gas chromatography with an electron-capture detection. The detection limits of cyanide and thiocyanate were 0.01 and 0.003 μmol/ml, respectively, while the gross recovery of both compounds was 80%. The calibration curve was linear over the concentration range from 0.02 to 1.0 μmol/ml for cyanide and from 0.01 to 1.0 μmol/ml for thiocyanate. The accuracy and precision of the method were evaluated, and the coefficients of variation were found to be within 10%. Using this method, the blood levels of two victims who had died from cyanide poisoning were determined.
AB - We devised a sensitive and simple method for determining cyanide and its major metabolite, thiocyanate, in blood using an extractive alkylation technique. Pentafluorobenzyl bromide was used as the alkylating agent, and tetradecyldimethylbenzylammonium chloride was used as the phase-transfer catalyst. The derivatives obtained were analyzed qualitatively by gas chromatography-mass spectrometry and quantitatively by gas chromatography with an electron-capture detection. The detection limits of cyanide and thiocyanate were 0.01 and 0.003 μmol/ml, respectively, while the gross recovery of both compounds was 80%. The calibration curve was linear over the concentration range from 0.02 to 1.0 μmol/ml for cyanide and from 0.01 to 1.0 μmol/ml for thiocyanate. The accuracy and precision of the method were evaluated, and the coefficients of variation were found to be within 10%. Using this method, the blood levels of two victims who had died from cyanide poisoning were determined.
UR - http://www.scopus.com/inward/record.url?scp=0030066613&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0030066613&partnerID=8YFLogxK
U2 - 10.1016/0378-4347(95)00344-4
DO - 10.1016/0378-4347(95)00344-4
M3 - Article
C2 - 8634765
AN - SCOPUS:0030066613
SN - 1572-6495
VL - 675
SP - 27
EP - 32
JO - Journal of Chromatography B: Biomedical Applications
JF - Journal of Chromatography B: Biomedical Applications
IS - 1
ER -