TY - JOUR
T1 - Comparison of two types of matrix-assisted laser desorption/ionization time-of-flight mass spectrometer for the identification and typing of clostridium difficile
AU - Kiyosuke, Makiko
AU - Kibe, Yasushi
AU - Oho, Megumi
AU - Kusaba, Koji
AU - Shimono, Nobuyuki
AU - Hotta, Taeko
AU - Kang, Dongchon
AU - Shoubuike, Takeo
AU - Miyamoto, Hiroshi
N1 - Publisher Copyright:
© 2015 MS Printed in Great Britain.
PY - 2015/10
Y1 - 2015/10
N2 - Microflex LT (Bruker Daltonics) and VITEK MS (bioMérieux) are bacterial identification systems that are based on matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). For VITEK MS, two identification softwares, VITEK MS IVD (IVD) and SARAMIS (SARAMIS), are available. Microflex LT is equipped with MALDI Biotyper RTC software (Biotyper). Although the identification accuracy of each instrument has been compared for various bacteria, no detailed examination has been conducted for the identification accuracy of Clostridium difficile. In this report, we compared the three identification softwares for identification reproducibility in three ATCC C. difficile strains and identification accuracy in 50 clinical C. difficile isolates. The results showed 100, 91.7 and 100% identification reproducibility accuracy of ATCC strains when examined by IVD, SARAMIS and Biotyper software, respectively. For the identification of the clinical isolates, all three softwares exhibited satisfactory identification accuracy of C. difficile. Among the 50 clinical isolates, seven showed identical toxin genotype corresponding to the exact ribotype. However, MALDI-TOF MS failed to identify them as the identical type. Based on the above results, we concluded that both types of MALDI-TOF MS reproducibly identified C. difficile; however, they are currently not suitable for typing of C. difficile clones.
AB - Microflex LT (Bruker Daltonics) and VITEK MS (bioMérieux) are bacterial identification systems that are based on matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). For VITEK MS, two identification softwares, VITEK MS IVD (IVD) and SARAMIS (SARAMIS), are available. Microflex LT is equipped with MALDI Biotyper RTC software (Biotyper). Although the identification accuracy of each instrument has been compared for various bacteria, no detailed examination has been conducted for the identification accuracy of Clostridium difficile. In this report, we compared the three identification softwares for identification reproducibility in three ATCC C. difficile strains and identification accuracy in 50 clinical C. difficile isolates. The results showed 100, 91.7 and 100% identification reproducibility accuracy of ATCC strains when examined by IVD, SARAMIS and Biotyper software, respectively. For the identification of the clinical isolates, all three softwares exhibited satisfactory identification accuracy of C. difficile. Among the 50 clinical isolates, seven showed identical toxin genotype corresponding to the exact ribotype. However, MALDI-TOF MS failed to identify them as the identical type. Based on the above results, we concluded that both types of MALDI-TOF MS reproducibly identified C. difficile; however, they are currently not suitable for typing of C. difficile clones.
UR - http://www.scopus.com/inward/record.url?scp=84944754866&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84944754866&partnerID=8YFLogxK
U2 - 10.1099/jmm.0.000136
DO - 10.1099/jmm.0.000136
M3 - Article
C2 - 26296999
AN - SCOPUS:84944754866
SN - 0022-2615
VL - 64
SP - 1144
EP - 1150
JO - Journal of Medical Microbiology
JF - Journal of Medical Microbiology
IS - 10
ER -