Analysis of sake mash using multichannel taste sensor

Yukihiko Arikawa, Kiyoshi Toko, Hidekazu Ikezaki, Yoshio Shinha, Teruo Ito, Isamu Oguri, Shigeru Baba

Research output: Contribution to journalArticlepeer-review

36 Citations (Scopus)

Abstract

The titrable acidity of and the ethanol concentration in sake mash were measured using a multichannel taste sensor system. A positively charged membrane responded to ethanol and the electric response of the membrane showed good agreement with those determined by gas chromatography in the measurement of sake mash. It was considered that the responses were induced by a change in the activity coefficient of anions. The titrable acidity of sake mash was measured using an electrode with a negatively charged membrane, because the electric response of the membrane showed n strong correlation with the titrable acidity of sake. The result showed good agreement with the value determined by titration, and the agreement was slightly increased by multiple regression together with the electric response of the positively charged membrane. Therefore, it is concluded that this method is useful for determination of the ethanol concentration in and the titrable acidity of sake mash. The results show that the taste sensor can be used for monitoring of those values in sake fermentation.

Original languageEnglish
Pages (from-to)371-376
Number of pages6
JournalJournal of Fermentation and Bioengineering
Volume82
Issue number4
DOIs
Publication statusPublished - 1996

All Science Journal Classification (ASJC) codes

  • Biotechnology
  • Applied Microbiology and Biotechnology

Fingerprint

Dive into the research topics of 'Analysis of sake mash using multichannel taste sensor'. Together they form a unique fingerprint.

Cite this