A simple method of energy calibration for thin plastic scintillator

S. Hohara, F. Saiho, J. Tanaka, S. Aoki, Y. Uozumi, M. Matoba

Research output: Contribution to conferencePaperpeer-review

1 Citation (Scopus)

Abstract

A thin plastic scintillator is used as a trigger counter of charged particle spectrometers. Since its Compton edge is not clear, it is difficult to check up the thin plastic scintillator counters with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured. The simulated spectrum was made with a Monte Carlo program considering geometries of the plastic scintillator. This technique was examined with the plastic scintillator having thickness of 2mm to 10mm.

Original languageEnglish
Pages6/226-6/230
Publication statusPublished - 2000
Event2000 IEEE Nuclear Science Symposium Conference Record - Lyon, France
Duration: Oct 15 2000Oct 20 2000

Other

Other2000 IEEE Nuclear Science Symposium Conference Record
Country/TerritoryFrance
CityLyon
Period10/15/0010/20/00

All Science Journal Classification (ASJC) codes

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'A simple method of energy calibration for thin plastic scintillator'. Together they form a unique fingerprint.

Cite this